Description
SPECIFICATIONS:
Defibrillator
- Operation: Fully automatic version.
- Energies: Standard max 200J or Power max 360J
- Waveform: Adaptive BTE (biphasic truncated exponential) conforming to patient chest's impedance.
- Protocols: Various adult shock protocols available on request.
- Factory default: Adult Standard escalating 150, 200, 200J. Adult Power escalating 200, 250, 360J, Pediatric (Standard or Power) 50J fixed.
- Charging time: <9seconds with a new and fully charged battery depleted battery will result in a longer charging time.
- Analysis time: IEC/EN 60601-2-4 from 4 to 15 seconds.
- Impedance: 200-200 ohms
- Sensitivity: IEC/EN 60601-2-4 (AHADB, MITDB source), 97%
- Specificity: IEC/EN 60601-2-4 (AHADB,MITDB source), 99%
- Controls: 2 buttons: ON/OFF, Shock, "i" info button.
- Flashing Icons: "connects pads to patient", "adult/child" informing on pads type use, "don't touch patient" warning to stay clear, "touch patient" informing it's safe to touch.
- Indicators: Status LED indicator informing on device condition. Battery gauge with remaining capacity rate. Audible alerts and text display with service alarms.
- Upgradable: Through a USB cable or memory card.
Battery Options
- Type: Li-SOCI2 Disposable, code SAV-CO903.
- Autonomy: 300 complete rescue cycles (shocks at 200J and CPR) or 200 complete rescue cycles (shocks at 360J and CPR) or 35 hours ECG analysis for a new and fully charged battery.
- Shelf-Life: When stored in original packaging, 5 years.
- Battery-Life: 4 years once installed to AED, assuming one battery insertion test and daily self-test but without switching AED on.
- Type: Li-ion Accumulator, Code SAV-C0011.
- Recharging time: 2.5 hours with the charger station Code SAV-C014. Recommended to charge every 4 months at least.
- Autonomy: 250 shocks at 200J or 160 shocks at 360J or 21 hours in ECG analysis for a new fully charged accumulator.
- Battery-Life 2 years or 300 charging cycles.
Pad Options
- Type: Disposable, pre-gelled and self-adhesive.
- Adult: Code SAV-C0846, for patient >8 years or >25kg.
- Pediatric: Code SAV- C0016, for patient <8 years or <25kgs.
- Cable length: 120 cm
- Shelf-Life: 30 months.
Event
Recording:
- Internal memory: up to 6 continuous hours of ECG and rescue events.
- Optional memory: Removable SD card; length of storage depends on card capacity: a 2GB card records up to 100 hours.
- Data recording: "AED1LOG" text file with detailed self-test activity. "AEDFILES" multimedia files with complete recorded information.
- Event review: "Saver View Express" data manager software.
Physical
- Size: 26.5 x 21.5 x 7.5cm
Environmental
- Operating temperature :0°C to 55°C (32°F TO 131°F)
- Storing/Shipping temperature: -40°C to 70°C (-40°F TO 158°F) without battery
- Humidity:10% to 95% relative humidity noncondensing
- Sealing (IP Protection): IEC/EN 60529 class IP54; splash proof, dust protected
- Shock/Drop Abuse Endurance: IEC/EN 60601-1 clause 21; 1 meter drop, impact, force, rough handling, mobile tolerance
- Electrostatic IEC/EN 61000-4-2
Discharge:
- Electromagnetic Compatibility: IEC/EN 60601-1-2 Emission, Immunity
- Electrical Protection: IEC/EN 60601-1; Internally Powered, Type BF
- Directive93/42/CEE and2007/47/CE: Class IIb (*)
- Temperature at20°C Humidity45% non-condensing